Abstract:
<p>{PURPOSE: To estimate the conversion factors to transpose macular thickness measurements on time-domain (TD) to various spectral-domain (SD) optical coherence tomography (OCT) machines in patients with macular telangiectasia type 2a (MacTel). PROCEDURES: Macular scans on TD- and SD-OCT were performed on patients at the same visit. The retinal thickness values in various ETDRS subfields and macular volume were compared between different OCT machines. RESULTS: The macular thickness and volume were significantly greater (p < 0.0001</p>